When the matrix and combustion behavior are unknown, there is no reliable universal sample weight that guarantees reliable elemental analysis results.
Start by considering four questions:
1)
What are you measuring?
CHNS, CHNS-O, carbon and nitrogen, TOC (Total Organic Carbon), or nitrogen/protein?
2)
How much material is available?
Limited material may require micro-scale analysis; abundant material gives more flexibility.
3)
How homogeneous is the sample?
Heterogeneous samples may require finer grinding or a larger aliquot.
4)
How does the matrix behave during combustion?
Organic-rich samples generate greater gas loads, while mineral-rich samples can often tolerate larger total masses.
5) What is the expected concentration of the target elements?
Samples containing low concentrations of the target elements may require larger aliquots to generate a sufficiently strong analytical signal, while highly concentrated samples may require smaller masses to remain within the detector's measurement range.
A practical way to establish the appropriate range is to compare results at different sample masses, provided the selected masses are within the analyzer's validated operating range. Agreement between different sample weights can help confirm that the combustion load is appropriate; divergence may indicate that the larger mass is not suitable or that another matrix-related issue needs to be investigated.
In short
The objective is to identify a "sample-weight window", rather than assume that one fixed mass will work for every unknown material.
For more information, see: [
Does a larger sample give better elemental analysis results?]