Why doesn't my CHNS elemental analysis match the calculated values?

When the experimental Carbon (C), Hydrogen (H), Nitrogen (N), and Sulfur (S) values of a compound do not match the calculated composition, several causes should be considered.

Possible contributors include:
  • Residual solvent or moisture;
  • Sample purity;
  • Incomplete combustion;
  • Weighing uncertainty;
  • Sample volatility;
  • Insufficient sample representativeness;
  • Analytical variation.
Weighing becomes particularly important at very low sample masses. With a balance offering 1 µg readability, for example, a 50 µg sample has a relative readability contribution of approximately 2%, while a 1 mg sample is approximately 0.1%.

For organic compounds, the commonly used ±0.4% difference between calculated and found values is a useful benchmark in many publications, but the appropriate acceptance criterion depends on the application and analytical context.

In short

When a result falls outside the expected range, it is therefore worth checking the sample itself, the preparation procedure, the weighing step, and the combustion conditions before assuming that the elemental analyzer is responsible.

For more information, see: [How should I prepare a sample for elemental analysis?] / [What is the minimum sample weight for CHNS elemental analysis?]


Find the Right Elemental Analyzer for Your Needs

Want to discover which elemental analyzer best suits your application?

Compare our solutions, explore their capabilities, and find the right fit for your sample types and analytical requirements!  
 
Find Your Ideal Analyzer
Contact data
Company data
Request details
I have read the privacy policy and accept the processing of data for the purposes indicated *
I agree to receive communications from VELP *
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Below all content tagged with: Why doesn't my CHNS elemental analysis match the calculated values?

Knowledge, News, Newsroom: Why doesn't my CHNS elemental analysis match the calculated values?